Evaluating a Novel AI Tool for Automated Measurement of the Aortic Root and Valve in Cardiac Magnetic Resonance Imaging

Parker J, Coey J, Alambrouk T, Lakey SM, Green T, Brown A, Maxwell I, Ripley DP. Evaluating a Novel AI Tool for Automated Measurement of the Aortic Root and Valve in Cardiac Magnetic Resonance Imaging. Cureus. 2024;16(5):e59647. doi:10.7759/cureus.59647.

This is research work and is separate from AortaAIM.

The study

The paper compared an AI tool with manual measurements on cardiac MRI from 35 patients in a registry, scanned on two 1.5T scanners. Patients with a bicuspid valve, known congenital heart disease or previous aortic surgery were excluded. Measurements included the aortic root area at the sinuses of Valsalva, the aortic valve area at the annulus in systole, and cusp-to-cusp and cusp-to-commissure distances in systole and diastole. Three NHS consultant cardiologists measured the same images as the reference.

Results

Agreement between all observers, including the AI tool, was high: an intraclass correlation of 0.959 in systole and 0.962 in diastole, against 0.972 and 0.969 for the cardiologists alone. The tool took 2.6 seconds per case; the cardiologists took 334.5 seconds on average.

Limitations

The authors note the small sample, the need for validation in larger and more varied groups and at other anatomical sites, and that effects on patient outcomes were not assessed. Several authors work for or hold interests in AIATELLA.